【概述Summarize】當溫濕度控制器顯示濕度異常的狀況,該如何辨別以及解決?
When the temperature humidity controller shows the condition of abnormal humidity, how ?
恒溫箱主要用于測試和確定電工、電子及其他產品及材料進行高溫、低溫、交變濕熱度或恒定試驗的溫度環境變化后的參數及性能。Thermostat is main used for testing and determining the parameters and performance of electrical and electronic and other products and materials after the temperature environment changes in high and low temperature alternating humidity and heat degree or constant test.
【檢查步驟 Check Steps】
查看控制機房的欠水超溫保護器,是否設定在150℃的位置。
Check whether the under-water and over-temperature protector in the control unit is set at the position of 150 Celsius.
查看機房內,管路是否過臟,造成水流不順。
Check the unit, whether the pipeline is too dirty, causing water flow.
查看上水箱是否有水,如果沒有水,再查看下水箱是否有水。
Check whether has water in top water tank, if not, same to check the bottom water tank.
查看是否濕度部份的固態繼電器燒毀短路: 如果加熱器未燒毀,使用三用電表交流電壓檔,電壓檔位開到600伏特的位置,將紅黑棒分別放在線號標注為H的那一顆固態繼電器交流兩側,將濕度部份的濕度設定值設置0%,此時濕度部份的固態繼電器的指示燈不會亮起,如果量測的電壓值沒有變化,維持在10V以下代表固態繼電器燒毀呈現短路狀態。或者首先將濕度部份設定值設為0%,再看機房內加濕桶是否水在煮沸狀態。
Check if the humidity of solid state relay burned short-circuit: if heater is not burning, used three meters ac voltage, the position of the gear drive voltage to 600 volts, will be red and black on the annotation of line number for H that a solid state relay communication on both sides, set the humidity of humidity value 0%, humidity at this time of the light on the solid state relay can't light up, if the measured voltage value did not change, keep under 10 v represents the solid state relay burned present state of short circuit Or the first set of 0% humidity part of the set value, then see whether humidifying barrels of water inside the room in the boiling state.
【解決辦法 Solutions】
1、將欠水超溫保護器轉至150℃的位置。
Turn under-water over-temperature protector to position 150 Celsius.
2、首先將機臺后方總電源關上,然后將加濕桶下方的排水閥,打開到與管路平行的位置,帶上水箱的水都排光的時候,利用板手將加濕桶打開,將內部因長期做濕度的水垢清除,然后將硅膠管路拔除,在水中搓洗。保養完畢后按照拆除的方式,反方向裝回去。裝置完畢后,需將加濕桶下方的排水閥,關閉到與管路垂直的位置,然后再將機臺后方總電源打開。
Firstly to turn off the total power supply behind the machine, then open the drainage valve under the humidifying barrel, till to a position parallel to the pipe line. Water shortage from top water tank, open the humidifying barrel by the spanner, scale removal and silica gel tube was removed, clean in water. After the maintenance in accordance with the demolition mode, put it back in the opposite direction, after installation, turn on the drainage valve with the pipeline on the vertical position, then turn on the total power supply.
3、下水箱都無水,將下水箱加水,即可解決。上水箱如果沒有水,下水箱有水,則為下水箱打水馬達壞掉,通知皓天黃工維修客服部門上門進行維修;
There is no water in the bottom tank, will add water in that tank, solve it ! If there is no water in the top tank, there is water in bottom tank, the bottom tank to drain water, means the motor broken. So inform the maintenance customer service department of Haotian company to conduct on-site.
4、更換濕度部份的固態繼電器(線號標注為H的那一顆固態繼電器),或者通知黃工售后上門進行維修。
Replace the solid state relay of the humidity part (the one marked with line No. H), or notify the after sales staff to conduct on-site maintenance.